North Temperate Lakes LTER Regional Survey Water Color Scans 2015 - current
Abstract
The Northern Highlands Lake District (NHLD) is one of the few regions in the world with periodic comprehensive water chemistry data from hundreds of lakes spanning almost a century. Birge and Juday directed the first comprehensive assessment of water chemistry in the NHLD, sampling more than 600 lakes in the 1920s and 30s. These surveys have been repeated by various agencies and we now have data from the 1920s (UW), 1960s (WDNR), 1970s (EPA), 1980s (EPA), 1990s (EPA), and 2000s (NTL). The 28 lakes sampled as part of the Regional Lake Survey have been sampled by at least four of these regional surveys including the 1920s Birge and Juday sampling efforts. These 28 lakes were selected to represent a gradient of landscape position and shoreline development, both of which are important factors influencing social and ecological dynamics of lakes in the NHLD. This long-term regional dataset will lead to a greater understanding of whether and how large-scale drivers such as climate change and variability, lakeshore residential development, introductions of invasive species, or forest management have altered regional water chemistry. Color is measured in water samples that are filtered in the field through 0.45 um nucleopore membrane filters. A spectrophotometer is used to quantify color in the lab as absorbance (unitless) at 1 nm intervals between the wavelengths of 200 and 800 nm. Absorbance data are considered suspect for values greater than 2.
Contact
Creator
Dataset ID
377
Date Range
-
LTER Keywords
Methods
We collect water samples for color at the deepest part of the lakes. The samples are surface water, filtered in the field through 0.45u polycarbonate membrane filters. We run a wavelength scan from 800 to 200nm, using a 5cm rectangular quartz cell in a Beckman Coulter Model DU800 spectrophotometer. Any samples that display absorbance values above 2AU are run again from 400 to 200nm using a 1cm quartz cuvette. Inititally the full range of wavelengths were run again and two values may be found in the database even if the original measurement with the large cuvette did not exceed 2AU. The user should discard values above 2AU and use values from the smaller cuvette instead. All values are given as measurements at the path lenth of the employed cuvette and need to be devided by the cuvette length for a comparable value at a pathlength of 1 cm.
The single beam Beckman Coulter DU800 spec is blanked first on a sample of DI water. Additional blank values are from a scan run on DI after that blanking as a check and are reported alongside the scans but are not subtracted from the scan values.
The single beam Beckman Coulter DU800 spec is blanked first on a sample of DI water. Additional blank values are from a scan run on DI after that blanking as a check and are reported alongside the scans but are not subtracted from the scan values.
NTL Keyword
NTL Themes
Version Number
3