Samples for chloride and sulfate are collected together with a peristaltic pump and tubing and in-line filtered (through a 0.40 micron polycarbonate filter) into new, 20 ml HDPE plastic containers with conical caps. The samples are stored refrigerated at 4 degrees Celsius until analysis, which should occur within 6 months. The samples are analyzed for chloride (and sulfate) simultaneously by Ion Chromatography, using a hydroxide eluent.
 
The detection limit for chloride is approximately 0.01 ppm and the analytical range for the method extends to 100 ppm.
 
Method Log: Prior to January 1998 samples, chloride was determined on a Dionex DX10 Ion Chromatograph, using a chemical fiber suppressor. From 1998 to 2011, chloride was determined by a Dionex model DX500, using an electro-chemical suppressor. From January 2011 until present, chloride is determined by a Dionex model ICS 2100 using an electro-chemical suppressor.

 