WSEL ICP
Instrument Operating Procedure
1) Turn
the chiller on, if it isn’t already on
2) Check
the Argon cylinder pressure.
3) Turn
on the computer and monitor
4) Click
on the WINLAB32 icon on the
desktop.
5) A
diagnostic window comes up with the System Status of “Spectrophotometer is in
standby mode”
6) Click
on the Spectrophotometer tab
and click on Initialize. Warm
up begins in 2 minutes and goes for another 75 minutes.
7) Replace
peristaltic pump tubing on the sample nebulizer assembly. Use red-red clear
tubing for the waste line and black-black solvflex tubing for the sample line.
Gradually stretch the tubing first before attaching the tubing to the
peristaltic pump to avoid uneven sections in the tubing when it is stretched
across the peristaltic pump rollers.
8) Click
Winlab32 Offline icon to
create a sample identification file while the spectrophotometer is still
warming up.
9) Click
on the file, then open sample info and select a
previous similar file or template. Modify the file with new sample ID’s, etc.,
and save the file as a new file with the project name and date of analysis. Ex.
LTER_042505.
10) Exit
Winlab32 offline, and place
the samples and standards in the autosampler tray, according to the sample info
file.
11) After warmup is complete, click on Plasma Control . Check the line
pressure, and if the pressure is less than 800kpa, slowly open the pressure
builder.
12) Click on Plasma On to ignite the plasma.
Shortly after the plasma lights, the peristaltic sample pump will start to
rotate.
13) Stretch the
pump tubing over the mounting brackets. Adjust the peristaltic line pressure
clamps to ˝ to 1 full turn from the end.
14) Observe the
nebulizer spray chamber, looking for a uniform and steady sample mist. The plasma should also appear steady
without flickers or pulses.
15) Click on tools, spectrophotometer control, and then Hg align. After the Hg wavelength alignment procedure is
completed, write the intensity count in the logbook.
16) Place the sample probe in a 1 ppm Mn solution.
Click align view and radial, to align the torch
position. After the torch has been aligned radially, click on axial to align
the torch along its axis, and write down the intensity in the logbook.
17) Click on file, open and method to select the method to be used. Choose the LTER
method that corresponds to the set of standards that are being run.
18) Select the
sample information file (that was just created offline) in the same manner as
step 17.
19) In the
sample information popup window, input the sample result file – usually the
analysis date.
20) Click
on Auto Shutdown and select At the end of analysis, place the
spectrophotometer on standby and turn the sample pump and plasma off after a 5
minute rinse.
21) After
the sample pump and plasma are on for over an hour, the system is stable enough
to start the analysis.
22) In the
automated analysis window select Run
Standards to have the instrument stop after calibration. Run Samples must then be selected
to start sample analysis. Select Run
All to run the samples automatically after running the standards
regardless of the quality of the calibration.
23) When the
analysis is done, cap the samples and standards as soon as possible to avoid
evaporation loss.
24) Turn the
water cooler off and close the argon tank pressure builder valve.\
25) Release the
sample pump tubing and clean the area, replacing printer paper and ink
cartridges as necessary.
26) Enter the
number of hours of ICP use (Argon use) in the logbook.
(updated 5/11/2005 jat)