WATER COLUMN AND SEDIMENT TRAP
PARTICULATE MATERIAL PROTOCOL
SUPPLIES NEEDED
PCTE filters with pore size of 0.4 um (Fisher
Catalog No. 09-732-37)
petri dishes for filter storage (Fisher Catalog No. 08-757-19)
microbalance with calibration weights
two flat-tip forceps
freezer storage space
oven capable of 40 - 60°C
in-line filter holders for TPM collection
sediment traps, covers, bottles and lines (minimum of 8 traps required for
LTER)
vacuum pump and filtration unit,
210mm mesh filter
centrifuge
freeze dryer
WEIGHING FILTERS
In the winter or early spring, weigh
450 filters for use in the summer. This
should be sufficient for all the TPM and sediment trap needs for the ice-free
season. Handle the filters by the edges
with a flat tipped forceps, placing each into a numbered petri
dish. Label five dishes as control
filters and the rest consecutively from YY-01 to YY-450, so that each
filter-dish combination has a unique identification number. Since the filters themselves cannot be
labeled, the filters must remain in the labeled petri
dish except when in use.
The day that filters are to be weighed, set the
sample filters and five control filters out on the counter by the balance, with
the petri dish lids partially open. Allow the filters to equilibrate with the
atmospheric humidity near the microbalance for several hours before
weighing. Cover the filters with a clean
plastic bag to prevent dust from settling onto them. Calibrate the microbalance before each use
using a 200mg weight. After every ten
filters weighed, check the tare, reset the tare and weigh a 10 mg weight to
determine if the microbalance is consistent.
If the 10 mg weight varies by greater than 5 micrograms from the initial
weight, recalibrate the balance. After
placing each filter on the balance pan, wait for the balance to give a constant
weight. This waiting period should be
standardized. We wait 45 seconds, then for the reading to be stable for 10
seconds before taking a number. Weigh
filters to the nearest microgram (e.g. 15.126 mg). Always engage the pan brake when loading or
unloading the balance. The stirrups that
the pans are suspended from are extremely fragile, and the balance is unusable
if they are broken off.
Differences in humidity between weighing dates
can greatly affect TPM values. The five control filters are used to correct for
humidity effects. These filters are weighed every time TPM or trap filters are
weighed. The average difference between the ‘before’ and ‘after’ weights of the
five control filters is used to correct sample filter weights for humidity.
At the end of the year, put the used TPM and
trap filters in a box with their control filters, and archive the box in a
dark, cool, dry area. Archived filters
may be analyzed for a variety of constituents.
Carbon cannot be determined, but particulate phosphorus, nitrogen, and
some metals can be extracted from the filters at a later date. Filters can also be examined under the
microscope using regular or Cyto-Clear slides.
TPM
SAMPLES
Total particulate material (TPM) is an estimate of the mass
of particles in the water column. In
this method particles are defined as material not passing through a
polycarbonate track-etched (PCTE) filter having a 0.4mm
pore size.
SAMPLE COLLECTION
Before going into the field, load preweighed filters into numbered inline filter holders,
recording filter number and filter holder on the datasheet. Collect TPM samples at the same depths as the
other LTER chemistry samples using an inline filtration setup. Filter directly into a graduated cylinder, pumping
until the flow is reduced to a drip or to 250 ml, whichever comes first. Record
the volume filtered on the field data sheet.
Remove and reattach the filter holder backwards on the line, then
reverse the peristaltic pump in order to suck the excess water through the
filter. Store the used filter holders in a cold cooler. At the lab, remove the
filters from the filter holders folding them in half with the sediment side in,
and return them to the numbered dishes.
Store them in the freezer until they can be dried.
PROCESSING SAMPLE FILTERS
Dry the filters in a 40-60°C
oven, in the petri dishes with the covers slightly
open, for 48 hours. After drying,
close the dishes and store the filters at room temperature until they can be
weighed. Follow the same weighing
procedure that is used for weighing filters before use.
CALCULATIONS
To determine the mass of particles in the water
column sample:
mass on filter (mg) / volume filtered (L) = TPM (mg/L)
The calculation is automated in the “wrkfile”
spreadsheet found on the
Trout lake server.
TPM data are then entered into the web based data form for inclusion in
the LTER database.
SEDIMENT
TRAP SAMPLES
The cylindrical traps are 44cm from the top of
the collection funnel to the top of the trap.
The collection area, calculated from the internal diameter of the trap
opening, is 165cm2 on the older Plexiglas traps and 185cm2 on
the newer pvc traps. The collection funnel inside the trap focuses
sedimenting particles into a detachable 250ml plastic
bottle. Total trap volume is 7420ml for
the pvc traps used beginning
in 2002. Before 2002, we measured the
volume of sample in the trap at the time of processing. The leak proof caps with the new traps
eliminated spillage during transport, so we now use a constant trap volume for
calculations.
TRAP AND LINE PREPARATION
Two mooring lines extend upward from an anchor
to a subsurface float wrapped in colorful tape.
The total length of the trap rig should be about 1m less than the depth
of the lake where the traps are deployed.
We use subsurface buoys to keep the mooring lines taut and to avoid
disturbance of the trap by passing boats.
Knots on the mooring lines to support the trap are placed so that the
mouth of the trap will be 7m off the bottom of the lake for Crystal
and Sparkling lakes and 8m off the bottom for Trout Lake. It is important that the two support knots
are exactly the same distance from the anchor or the trap will not hang
perfectly vertically in the water.
Place approximately 0.2g NaN3 into a
4ml vial with septum cap. NaN3
is a toxic substance and should be handled accordingly: read the cautions on
the bottle and wear gloves. Perforate
the septum several times with a hypodermic needle, fill the vial with milliQ water and drop it into the collection bottle. Fill the collection bottle with milliQ water and attach to the threaded collar on the
bottom of the trap.
TRAP DEPLOYMENT AND COLLECTION
Duplicate traps are deployed near the LTER
sampling buoy in Crystal, Sparkling and Trout Lakes. They should be placed at least 5m from each
other and at least 10m to the north of the sampling buoy. It is critical that they are far enough apart
and especially that they are far from the permanent anchor. If they are too close they become entangled
with one another or the permanent anchor and can become impossible to retrieve.
In the field, slip the mooring lines through the
slots on the base of the trap so that the knots are on the inside of the slots
supporting the weight of the trap. Run
the lines up the side of the trap and fasten into the clips near the top of the
trap. Lower the anchor into the water,
holding on to the mooring lines above the trap.
Push the trap into the water and allow it to fill, holding it down on
the mooring line knots until it is full.
If the trap is released without being filled, it will float off the knots
as it is lowered and will not hang vertically in the water column. When the trap is full, lower it into the
water adjusting the length of the mooring line if necessary so the float is
about 1m below the surface.
Sediment traps are left in the lake for
approximately one month, from one chemistry sampling date to the next. Retrieve the traps after all other sampling
has been completed. Snag the mooring
line with a hooked pole, pulling the line into the boat until the trap reaches
the surface. Cap the trap and remove it
from the lines, fastening it into the rack in the boat for transport. Replace with a fresh trap. Upon returning to the lab, process the traps
as soon as possible. Do not allow them
to sit for an extended period of time exposed to light and heat.
SAMPLE PROCESSING
We collect two sample bottles from each trap:
the collection bottle attached to the funnel, and a subsample
of the above-funnel water. Seal the
funnel opening with the long handled cork, and remove the collection bottle from
the bottom of the trap. Place the trap
in a plastic washbasin and remove the cork to drain the trap into the
basin. Stir the water in the basin and
scoop a subsample into a clean 250ml trap
bottle. The remaining water in the basin
can be discarded. The two traps from
each lake are arbitrarily designated Trap A and Trap B. Label the collection bottle and the
above-funnel bottle from each trap with a letter designation and the lake
name. Refrigerate until further
processing.
Rinse the traps with tap water, scrubbing with a
soft brush if necessary to remove any particles remaining on the sides of the
trap. Store traps clean and dry.
Final processing of the sediment trap samples
should be done as soon as possible after collection. Material in the collection bottle is split
into two size fractions by pouring it through a 210mm nylon mesh circle fitted into a filter funnel. The 210mm separation removes the extra variability in sedimentation rate
due to large zooplankton and large detritus that may be found in a sediment
trap. There are chironomids
in traps from most lakes, and Mysis in Trout Lake
traps. Since these things are not really settled material (they migrate in and
die), we want to remove them. To
facilitate filtration, milliQ water may be squirted
into the filter funnel to loosen particles covering the mesh, and very gentle
suction may be applied. Rinse the sample
bottle with milliQ water and pour the rinse water
through the mesh.
Weigh the twelve plastic bottles and record the
empty weight in Column A on the data form.
Pour the sample fraction that passed through the mesh into the
appropriate 500ml bottle. Wash the
material retained on the mesh through a glass funnel into the 250ml bottle with
milliQ water.
Weigh the sample bottles with sample and record in Column B on the data
form. Also weigh and record the weight
of the above-funnel bottle with sample.
Filter subsamples of each of the
three samples (<210mm, >210mm,
and above-funnel) through preweighed 0.4mm
pore PCTE filters using suction <15 psi. Shake the sample bottle prior to pouring so
that the subsample filtered is representative of the
entire sample. In general you can only
filter 10-20ml of the <210 sample, 40-60ml of the >210 and ~100ml of the
above-funnel sample before the filter is clogged. As long as particles are visible on the
filter, there is enough to weigh. Do not
filter too much because if the mass on the filter is too great, it will flake
off after drying causing a loss of particulate matter. Return each filter to its numbered petri dish, folded in half with the sediment side in. Freeze the filters until they can be dried
and weighed. Record the number of the
filter used in Column D on the data form.
Weigh the three sample bottles after filtering and record the weight in
Column C on the form.
Discard the remaining
above-funnel sample. Centrifuge the
entire remainder of the <210 and >210 samples at 7000rpm for 30 minutes
at 4°C. [Note: at Trout Lake we do not have a chilled centrifuge or one that spins
at 7000rpm. Use 3200rpm for 30-50
minutes.] Decant and discard the
supernatant and place the muck pellet in a small, labeled plastic vial and
freeze. Wash all trap bottles with milliQ water, allow to dry, and reuse
them. Freeze dry the centrifuged
samples, and store in a cool, dry place.
CALCULATIONS
To determine the deposition of particles into the lake for
each of the three fractions (<210mm, >210mm,
above-funnel):
mass on filter
(mg) / volume filtered (ml) = particle concentration in subsample
(mg/ml)
particle
concentration (mg/ml) * total sample volume (ml) = total particle
mass in sample (mg)
total particle mass /
area of trap mouth / length of deployment = particle mass flux (mg/cm2/day)
These calculations are automated in the “wrkfile”
spreadsheet found on the Trout
Lake
server. Add the fluxes of the three
fractions together to determine the total flux of particulate matter (mg
particles / cm2 / day) for each trap deployment.
(reviewed
2/05 pkm)