pH
AND ALKALINITY MEASUREMENT
EQUIPMENT
Orion 720A pH Meter
Radiometer Model GK2401C Combination
Electrode
SAMPLE
COLLECTION
Rinse bottle 3 times with small volumes of
the water to be sampled. Collect pH and alkalinity
samples in separate 20-ml scintillation vials with displacement caps. Collect a second pH sample in a 125-ml wide
mouth bottle. Fill the 20-ml vials with
a minimum of splashing and air entrainment.
Fill to overflowing and cap, excluding all air bubbles from the sample. It is important to ensure that no atmospheric
gas exchange occurs between the time of sample collection and analysis. The 125-ml bottles are for air equilibrated
pH determination, and may have air in them.
Keep samples cold and dark to minimize biological activity.
pH
MEASUREMENT
pH samples should be analyzed as soon as
possible after collection, and must be analyzed the same day they are
collected.
Warm samples to room temperature in a dark
container. The container may be filled
with room temperature water to speed the warming.
Fill the electrode syringe barrel with an
extra sample of the lake water to be analyzed.
Leave the electrode to soak in this sample at least 15 minutes. The longer the electrode is
"conditioned" before pH measurement, the faster it will respond
during measurement.
Bubble air through the samples in 125-ml
bottles for at least 15 minutes before analysis using a vacuum pump and the
gang valve assembly. The intake hose
should be pulling air from outside the building. After bubbling, cap samples tightly until
analysis.
Uncover the
electrode filling solution hole. Remove
the conditioning solution from the electrode syringe barrel using the three way
valve and aspiration system.
With a clean 5-ml syringe take an aliquot of
the sample to be measured and inject about 2 ml into the electrode
chamber. Drain chamber. Inject another 2 ml and flush in and out of
the chamber several times. Drain.
Refill syringe with another aliquot of the
sample and inject 2 ml into the electrode chamber. There should be enough sample in the chamber
to cover the electrode junction.
Press the MEASURE button on the pH meter to
take the meter off standby. The meter
should be in millivolt mode; if not, press MODE until the first line of the
display indicates MV. The second line of
the display should read: CH-1 25.0 C
MEAS. If it reads HLD rather
than MEAS, press the MEASURE button again to begin measurement.
When the electrode response is stable the
meter will beep and indicate HLD on the display. Record the millivolt value on the data sheet.
Continue to run 2-ml aliquots of the same
sample until two consecutive millivolt readings are within 1 millivolt. Record the last millivolt value in the
FINAL MV column of the data sheet.
When all samples
have been analyzed, obtain mV readings for the 3 buffer solutions: pH 10.00, 7.00, and 4.00. Only one 2-ml aliquot of each buffer need be
run.
Rinse the
electrode chamber with Trout Lake tap water and store the electrode in Trout Lake tap water. Cover the
electrode filling solution hole securely with tape. Rinse all sample bottles and syringes three
times with milliQ water and leave out to dry.
Leave the pH meter on standby.
(reviewed
2/05 pkm)
ALKALINITY MEASUREMENT
Note: This Alkalinity protocol which is used for analyses
done at both the Trout Lake Station and the Center for Limnology. The deviations
from the standard protocol for the Center for Limnology are shown in parenthesis
below):
Warm
samples to room temperature in a dark container.
Place the electrode in a sample of water
from the lake that the alkalinity samples are from. Allow to soak for at least 15 minutes.
Turn the stirrer on 15 minutes before
beginning titrations to allow it to come to constant stirring speed. Do not change the stirring speed during
titrations and buffers. A change in
stirring rate introduces a millivoltage error called the 'streaming potential'.
Fill the Gilmont buret with standard 0.01N
HCl as follows (0.05N HC for CFL): Place
the delivery tip of the buret in a waste beaker and deliver acid until there
are no air bubbles in the tubing. Rinse
with milli-Q water, wipe dry, and deliver a few more drops of acid. Place the tip in a beaker of standard HCl and
draw acid into the buret. Rinse and
wipe tip again, and deliver acid into the waste beaker until the buret is
zeroed. If there is air in the buret,
remove it from the clamp, invert, and deliver the air out the tip.
Run a 'dummy' alkalinity titration on a
sample of conditioning water from the lake before running any samples. This is done because the first alkalinity
titration often gives a suspicious value.
Rinse
a small stir bar with milli-Q water, dry with a kimwipe, and place in a clean,
dry 2-dram vial (25 ml vial for CFL). Do
not touch the stir bar with your fingers; handle with a magnet and forceps
only.
Place
3 or 4 ml of sample (16 ml of sample for CFL), depending on the expected alkalinity
of the sample, into the vial using an Eppendorf pipet. Draw up and deliver sample from the pipet slowly
so that no air bubbles are drawn into the tip, or sample droplets splashed
onto the side of the vial. Because
of the small volume titrated, any loss of sample by spattering or any sample
forming a droplet on the side of the vial will not be titrated, thus giving
low values. If you make an error, start
over with a clean, dry vial.
Remove
the electrode from the conditioning lake water vial. Rinse the electrode with milli-Q water and
carefully blot it dry with a kimwipe. Place
the electrode in the vial containing the sample, slide the magnetic stirrer
below the vial, and center it on the 'x' on top of the stirrer.
Zero
the acid buret, wipe the delivery tip, and insert it into the vial.
Note
the initial millivolt reading and add small increments of acid until the meter
reads >115 mV. When the electrode
response is stable, the meter will beep and the display will indicate HLD. Record the volume of acid added and the mV
reading as your first increment.
Titrate the sample from +115 to
approximately 165 mV using 10 equal increments of acid. Generally 0.010 ml increments are
required. Record the volume of acid
added and the equilibrium millivolt value after each increment.
At the end of the titration, place the
meter on standby. Remove the buret
delivery tip from the vial, rinse with milli-Q water, and dry. Deliver some excess acid from the tip to
flush the tip. Remove the stir bar from
the sample with the magnet, rinse with milli-Q water, and dry.
After completing sample titrations,
standardize the electrode using fresh buffer solutions (pH 4.00 and 7.00). Place a few mls of buffer and a stir bar in a
vial, and take a mV reading. Do not
place the acid buret tip in the buffer vials.
Turn off magnetic stirrer, place the pH
meter on standby, cover electrode filling solution hole. Rinse electrode and store in Trout Lake tap water. Rinse
sample bottles and titration vials three times with milliQ water and leave out
to dry.
(reviewed
2/05 pkm)