Method for
the analysis of anions using the Dionex DX500
Start up
and run
1.
Start up computer if not already on. Check the
paper supply in the printer.
2.
Check level of eluent.
For an overnight run of about 60 samples the level should be around 1600 ml. -
If low prepare more.
3.
Turn
on Helium gas at the top of the tank. Do not adjust the line pressure.
4.
Open
Peaknet main menu.
It is located in the start menu under “Programs/Dionex
Programs”.
5.
Degas
the eluent. To do this place the plastic tubing in
the helium outlet on the bottom of the cap and place tube in the eluent. lightly screw on the cap a few turns. Make sure the cap is not sealed tight for the degassing procedure. Check to see if there is a Wipe-All in the
tray beneath the bottle to contain any spill.
Degass for about 5 to ten minutes. When finished remove the plastic tubing that
was installed and replace cap, sealing it this time.
6.
If
new eluent was added prime the pump. This is done by first loading a method to
start the pump. To load a method, open
the RUN window in the Peaknet program. Under “file” on the menu bar select “load
method”. Select the method you wish to
run. In most instances it will be
“anions.” This will start the pump. Under “run” on the menu bar select direct
control of the GP50. Highlight the low
pressure limit and change to “0.” Next open the bottom
door to the GP50 and locate the pressure transducer valve.(top center on the
pump.) Turn this valve one or two turns
counter clockwise. On the direct control
screen press the prime button. This
will start the pump priming. let it continue until all air is expelled from the
tubing. Check the sample waste line for
a reference. This should only take about
one or two minutes. When it clear, press
the prime button again to stop the priming process. close the pressure
transducer valve. In the direct control
screen re-enter the low pressure limit for the pump(1000).
7.
If
not already loaded, load the method you plan to use for the day. this is done by
first entering the RUN screen from the peaknet main
menu. under
“File” select “load method” Load the “anions” method.
8.
With
the method loaded, we must establish a base line to determine stability. Under the “run” menu select “baseline”.
9.
Prepare
a schedual for today’s run. Important, create a new schedual for each
run do not modify and old schedual. Enter the SCHEDUAL editor from
the peaknet main menu. Under “file” select “new schedual”. Type in the standard and sample ID’s. Identify the entry as sample, calibration standard
or check standard. Identify the level of
each of the Calibration and Check Standards.
Double click on the box to enter the method. select a method for
the first entry. Use the fill button
located on the menu bar to aply that method to the
rest of the samples. Enter a datafile name for the first entry (Use the date for easy
reference) and fill the rest using the fill button, just as with the method. At the end of the schedual
enter an ID called shutdown. Identify
the shutdown method for this entry.
Enter a file name to avoid a error message when
saving the schedual.
Save the completed schedual using the date as
the file name. this will aid in the future reference
of data files.
10.
Load
the auto sampler with the samples. The
black dots on the shuttles should go tho the
right. The caps should be aproximately flush with the top of the vial. To advance to the first sample press the hold
button on the front of the autosampler and it will
advance to the sample.
11.
In
the RUN screen from the Peaknet main menu, under “file”
select “Load Schedual”. Under the “run” menu select “Start”. To avoid complications in the processing of
the final data be sure that the Schedual is completed
and saved in it’s final form before starting the run.
Tell the machine to run a batch file at the end of the run. Identify the file to be named the date of the
run.
Shutdown Procedure
1.
After the last sample is finished, check to
see if a shutdown method was run. If so
the supressor column and the pump will be off. If not enter the RUN screen from the Peaknet main menu.
Under “file” select “Load method”.
Load the “Shutdown” method.
2.
To
see if the supressor is off view the SRS reading on
the main screen of the ED40. On the GP50
the ON/OFF button will higlight off when it is not
running.
3.
Turn
off the gas at the top of the tank. Do
not adjust the line pressure.
4.
If
a batch process was not run at the end of the schedual,
perform a batch process on the data.
5.
Flush
out the back of the pistons.
6.
If
leaving over the weekend Shutdown the computer.
This is done from the start menu.
It may also be wise to turn off the DX500.