CONDUCTIVITY
Equipment: -YSI
Model 32 Conductance Meter
-YSI
3403 Conductivity Cell
(Cell Constant = 1.0)
-YSI
3220 Temperature Sensor
Preparation: -Bring
all samples and standards to 25 C (+/- 1 C) in a water bath.
-In
addition to the lake water samples, you should have standards of 1400 µmho/cm, 10
µmho/cm, and 100 µmho/cm, and samples of water from the Trout Lake Station
water purification systems.
-Connect
the conductivity cell leads to the terminals marked "CELL".
(The lead ends may go on either terminal.)
-Plug
the temperature probe into the proper socket.
1. Determine the cell
constant:
-Shake
the bottle of 1400 µmho/cm standard to ensure solution homogeneity.
-Rinse
the plastic sample tube, the conductivity cell, and the temperature probe twice
with the standard.
-Fill
sample tube and place the conductivity cell in sample, making sure that the
vent slots are submerged and there are no air bubbles trapped inside the
conductivity cell.
-Dial
the FUNCTION knob to 'TEMPERATURE' and read the sample temperature to a tenth
of a degree. Allow enough time for the
meter to stabilize before recording the temperature.
-Dial
the FUNCTION knob to the first 'CONDUCTANCE' position and record the sample
conductivity. Conductivity should be
read on the lowest range possible. If
the conductivity reading is followed by a 'u', switch to a higher range.
-Take
temperature and conductivity readings on three more aliquots of the same
standard.
-Calculate
the 'CA' value for each aliquot, and use the average CA to calculate the cell
constant. Report the cell constant to
.001. The cell constant should be
between 0.98 and 1.02.
-Rinse
the sample tube, conductivity cell, and temperature sensor three times with milli-Q after running the standard.
2. Measure the
conductivity of the remaining standards and samples:
-Run
the station RO and deionized water samples first,
followed by the 10 µmho/cm standard, followed by the samples in order of
increasing conductivity. Run the 100
µmho/cm standard last.
-Shake
each sample and standard before measuring conductivity.
-Rinse
the sample tube, conductivity cell, and temperature probe with two aliquots of the
standard or sample to be measured.
-Place
conductivity cell in sample, and record temperature and conductivity in the
same manner as for the 1400 µmho/cm standard
-Pour
a second aliquot and record temperature and conductivity as for the first. If the second conductivity value differs from
the first by more than 0.5 µmho/cm, continue to run additional sample aliquots
until two consecutive readings are within 0.5 µmho/cm of each other.
-Always
run three aliquots of the standards and purified water samples even if the
first two are in agreement.
3. Calculate corrected conductivity values from
the measured temp. and cond. values:
-Use
the conductivity calculation worksheet on the Station server to calculate final
conductivity values. Fill in the
measured temperature and conductivity values.
The 'average conductivity' column contains the values that will be
entered into the LTER database.
4. Clean-up:
-Turn
off the meter and store it in the chemistry lab.
-Rinse
the conductivity cell well with milli-Q and store it
in milli-Q.
-Rinse
sample and standard bottles three times with milli-Q. Place bottles in the hood or on drying rack
in the chemistry lab to dry
5. Additional
Comments:
-Samples
should be stored in the refrigerator before analysis. They should contain enough air space to allow
for mixing.
-Conductance
of some samples may drift, especially those from the bog lakes. In this case, the procedure is to measure the
sample as quickly as possible after shaking and pouring, and to record the
conductivity immediately after agitating the sample at the cell.
-The
coating of platinum black on the electrodes of the conductivity cell should be
inspected before and after use for signs of flaking or loss of material. If the coating seems poor, cleaning and
replatinization is required as described in the manufacturer's instructions.
EQUATIONS FOR CALCULATING CELL
CONSTANT AND SAMPLE CONDUCTIVITY
Conductivity is routinely reported as a value corrected to
25 C. This correction is represented in
the following equations as: (CM)(25 -
T)(0.0191)
where: CM =
measured conductivity
T
= measured temperature (C)
Calculation of Cell Constant:
K = ST/CA where: K
= cell constant
ST
= true conductivity of standard
CA
= calculated conductivity of standard
CA = CM+(CM)(25-T)(0.0191)
Calculation of Sample Conductivity: CA = K{CM+(CM)(25-T)(0.0191)}
(reviewed 1/05 pkm)