Daily
Operation of Technicon Autoanalyzer II
LTER-Geochemistry Lab
Start up:
NOTE: To save time, start a warm water
bath to bring the reagents up to room temp. Takes about 15
minutes to get hot water and about 20 minutes to warm the reagents.
1) Hook up the appropriate analysis
module to the colorimeter and the autosampler lines. Hook up bubble lines to
air pumping lines on the peristaltic pump. Check the pump tubing diagrams in
the Lab Methods Manual if you have any questions.
2) Place the appropriate pair of
filters in the colorimeter. Only the colorimeter in the back will ever need to
have the filters changed. The Nitrite/Nitrate module uses the 550 nm filter for
both Nitrite/Nitrate and Total Nitrogen analyses. The Silica module uses the
660nm filters; the ammonia module uses the 630 nm filters; the phosphorus
module uses the 880 nm filters.
3) Fill the Milli-Q reservoir 1/2 full
with fresh Milli-Q water.
4) Check the waste carboy. Empty if it
is 2/3 full.
5) Place the pump tubing reagent lines
into MQ while the reagents are warming up. Place the sampler(probe)
into the autosampler and let it run on MQ. Once the reagents have warmed up,
move the lines from the MQ tothe appropriate reagent bottles (lines and bottles
are labeled).
6) Turn on power switches at the outlet
board. Be sure the autosampler's power button is not engaged.
7) Install the platen plate on the
peristaltic pump assembly and run pump on high speed once to get reagents
flowing through lines.
8) Check to see if all reagents are
flowing smoothly through the pump lines. If not, change lines
position on the pump rollers or replace worn pump tubing.
9) After the bubbles in the lines are
flowing evenly and are uniform in size, turn on the colorimeter and plug in the
heating element(when needed). Set STD. CAL. knob to value used on previous analyses.
10) After 15 minutes, turn on the strip
chart recorder(s), and install the pen(s).
11) Use the baseline knob on the
colorimeter to establish a zero baseline.
12) Once a stable baseline has been established(usually 45-60 minutes), place the prepared tray
on the autosampler (a standard 1 and a Milli-Q sample should be run first for
calibration purposes). Press down the autosampler power button.
13) When the calibration standard's
peak comes through on the chart paper, adjust the STD. CAL. knob on the
colorimeter to attain a peak height that is almost the full width of the chart
paper at the peak's maxima. When the Milli-Q sample is being recorded, adjust
the baseline knob of the colorimeter to place the recorder pen at its original
zero position.
14) You now have the system calibrated
and should not need to adjust the STD. CAL. knob the remainder of the analysis.
If the baseline drifts up or down significantly over the course of the
analysis, you may adjust it with the baseline knob when a Milli-Q sample is
being detected. Record the adjustment with an arrow on the strip chart and
label it with a message.
15) A full tray takes one hour and
twenty minutes to be sampled completely. To assure that the autosampler stops
at the end of the tray, place the red peg in the hole by the second-to-last
sample cup. Continuous analysis may be maintained by taking the peg out and
being sure to replace the sampled tray with an unsampled tray after the last
sample has been aspirated.
16) Label the strip chart(s) with the
name of the analysis, date, and analyst.
Shut down:
(Takes
30-40 minutes at end of analysis.)
0) Remove
the cadmium column from the nitrate module without introducing any air into the
column.
1) Remove reagent lines to
a beaker filled with clean Milli-Q water. Run the pump on high speed one time.
Place reagent bottles back in the walk-in cooler.
2) Remove the reagent lines
to the small bottle half-filled with 0.1 N NaOH solution.
Run the pump on high speed one time.
3) Remove the reagent lines
to the beaker filled with Milli-Q water and run the pump one time on high
speed.
4) Remove
the reagent lines to a small bottle half-full of 0.1 N HCl. Run the pump on
high speed one time.
5) Remove the reagent lines
to the beaker of Milli-Q water and run the pump two or more times on high
speed.
6) Remove the reagent lines
to a plastic storage bag. Lift the autosampler "probe" out of the
wash receptacle. Run the pump on high speed until liquid is no longer visible
in the glass coils or the colorimeter tubing.
7) Remove the platen plate
assembly and use a Kimwipe to clean the gray underside with isopropyl alcohol.
8) Turn off the power
button on the autosampler and shut off all outlet power switches.
9) Turn off the power on
the chart recorder and cap the recorder pen.
10) Check the waste carboy,
empty if 2/3 full.
Additional Notes:
If you are having problems
with an unsteady or unusual baseline, the pump tubing and/or the pumping action
are suspect. Most problems encountered are due to irregularities in the pumping
action from worn pump tubing or blockages/air leaks in the lines.
A level but noisy baseline
may be caused by a small air bubble caught in the colorimeter's detection
pathway. This may be "pulled through" by squeezing for a few seconds
on the tube labelled TO PUMP and then releasing. A noisy baseline may also
indicate worn pump tubing or a reagent which is precipitating.
If you notice that liquid
is dripping from the connecting lines of the mixing module and the colorimeter,
you have probably forgotten to add a detergent (Aerosol 22 or Brij 35)to your reagents. See the Lab Methods Manual for details.
Tray Preparation for NO2&NO3/NH4
and ∑N/∑P Analyses:
1) Fill tray with
conical-bottomed cups
2) Fill
each cup completely with 1 N HCl. (83 mls conc. HCl/L)
3) Aspirate the acid from
each cup.
4) Fill each cup completely
with fresh Milli-Q water.
5) Aspirate the Milli-Q
water from each cup.
6) Repeat steps 4 and 5.
7) Aspirate thoroughly any
water droplets in the bottom or clinging to the sides of the cups. This step is
very important!
8) Fill sample tray
immediately with the samples.
Tray Preparation for
BRSi and DRSi analysis:
1) Fill tray with
flat-bottomed cups.
2) Fill the sample tray
with samples. (No washes or rinses are needed!)
last revision: 8/24/95 by James Thoyre